1999 Quality Assurance Project Plan for the Regional Monitoring Program
for Trace Substances
|
| Prepared by
Sarah Lowe, Rainer Hoenicke, and Jay Davis
San Francisco Estuary Institute
With contributions from
Genine Scelfo
University of California, Santa Cruz
Prepared for the
San Francisco Estuary Regional Monitoring Program
San Francisco Estuary Institute
2nd Floor
7770 Pardee Lane
Oakland, CA 94621
May 1999
RMP Contribution #33
|
Acknowledgments
We are grateful to the following individuals who have assisted in the
development of our quality assurance and control program as well as this
document.
Andy Gunther, Applied Marine Sciences
Genine Scelfo, University of California, Santa Cruz--Trace Metals Laboratory
Walter Jarman, University of California, Santa Cruz--Trace Organics
Laboratory
back to contents
|
Contents
Acknowledgments
1. Introduction
Definition of Quality Assurance and Control
2. Overview of the RMP
RMP Organization
3. Overview of Quality Assurance and Control in the RMP
Sample Collection, Preservation and Holding
Laboratory Operations
Information Management
Sample Tracking
Data Reporting Requirements
4. Field Quality Assurance and Quality Control
Field Performance Measurements: Terminology
Field Performance Measurements Used by
the RMP
5. Laboratory Quality Assurance and Control
RMP Laboratory Requirements
Laboratory Performance Measurements
Laboratory Quality Control Procedures
QA Procedures for Ancillary Parameters
Water
Sediment
Bivalves and Fish Tissue
References
Tables
Table 1
Table 2
Table 3
Table 4
Table 5
Download PDF version
|
1. Introduction
This document presents the San Francisco Estuary Institutes (SFEI)
quality assurance and quality control (QA/QC) protocols and requirements
for contract laboratories associated with the Regional Monitoring Program
for Trace Substances (RMP). It includes:
- A summary of the RMP and its organization.
- An overview of quality assurance and control in the RMP.
- Quality assurance and control measures in the field.
- Quality assurance and control measures in the laboratory.
Much of the guidance provided in this document is based on protocols
developed for the Bay Protection and Toxic Cleanup Program (BPTCP), EPAs
Puget Sound Estuary Program (US EPA, 1989), as well as those developed
over many years for the National Oceanic and Atmospheric Administrations
(NOAA) National Status and Trends (NS&T) Program. Many other individual
research and monitoring programs also provided guidance for this document.
Detailed descriptions of field and laboratory methods are available through
SFEI.
back to
contents
Definition of Quality Assurance and Control
Ideally, a monitoring program is based on specific management questions
which lead to the formulation of quantitative measurement endpoints. These
measurement endpoints are used to develop data quality criteria (DQCs)
and performance standards based on realistic confidence and certainty
levels. The analysis of monitoring samples requires specific guidance
from policy makers and environmental managers as to what the desired uses
of the data are. Conversely, what kinds of environmental management decisions
can be made in a scientifically defensible way depends on the sensitivity
of the measurement system and the levels of confidence and certainty in
the data. The purpose of this document is to maximize the probability
that environmental data collected by the RMP will meet the expectations
of the data users. The DQCs outlined in this document are intended to
ensure, to the greatest extent possible, that the data truly represent
conditions in the environment with negligible artifacts due to sample
collection and processing.
The RMP quality assurance and control system was designed to accommodate
evolving information needs by the data users within the inherent constraints
of the best available sampling and analytical methodologies. The acceptable
or unavoidable variability that is introduced through the sampling and
measurement system, as well as the desired sensitivity levels that allow
quantitative comparisons to receiving water quality objectives, are reflected
in the RMP DQCs expressed in terms of accuracy, precision, completeness,
and method detection limit requirements. The DQCs for the RMP were established
based on instrument manufacturers specifications, scientific experience,
and historical data. Individual contract laboratories are given the greatest
degree of flexibility in their analytical procedures, as long as they
can demonstrate that DQCs are being met and that data comparability between
laboratories and analytical matrices are documented.
Quality control can also be described as a system that accounts for and
quantifies as many potential measurement errors as possible in order to
evaluate the uncertainties associated with any given measurement. Errors
that influence environmental measurements can be introduced in the field,
during shipment, and in the laboratory. The following are some examples
of sources of field and laboratory contamination that may need to be taken
into account when evaluating sample data quality:
A. Field
- Sample containers
- Sample equipment (tubing, pumps)
- Ship (exhaust, metal surfaces)
- Personnel (dirty hands, general carelessness)
- Atmospheric deposition
- Preservatives
B. Laboratory
- Atmospheric deposition
- Personnel
- Chemical contamination from extraction and/or preparation
- Analytical instruments and equipment (tubing, corrosion, etc.)
- Reagents
- Containers
back to contents
|
|
2. Overview of the RMP
RMP Organization
Project Information
The Regional Monitoring Program for Trace Substances (RMP) began in 1993
and evolved out of a pilot program funded under the States Bay Protection
and Toxic Cleanup Program, after the San Francisco Bay Regional Water
Quality Control Board (Regional Board) had developed a funding and implementation
structure.
At this time, 74 public and private entities that discharge treated wastewater
and cooling water, that manage stormwater runoff, or that are involved
in dredging activities contribute the financial resources necessary to
conduct the RMP. Many of these program participants also contribute expertise
or logistical support. The San Francisco Estuary Institute (SFEI), as
the entity designated to implement the Regional Monitoring Strategy, is
administering the program under a Memorandum of Understanding with the
Regional Board.
Currently, about 170 individual chemical parameters are analyzed in water,
sediment, and tissue two to three times per year (Table
1). Bioassays on water and sediment samples are also conducted to
determine possible toxicity to selected organisms. Originally, most of
the station locations were chosen so they would be as far as possible
from the influence of major contaminant sources and to be as representative
as possible of "background" contaminant concentrations. In subsequent
years, more stations have been added that are located close to tributaries.
Two stations adjacent to the wastewater outfalls of the Cities of San
Jose and Sunnyvale are monitored using RMP methodology under a special
agreement with the two National Pollutant Discharge Elimination System
permit holders and the Regional Board.
back to contents
Objectives
The current objectives for the RMP are to:
- Describe patterns and trends in contaminant concentration and distribution.
- Describe general sources and loadings of contamination to the Estuary.
- Measure contaminant effect on selected parts of the Estuary ecosystem.
- Compare monitoring information to relevant water quality objectives
and other guidelines.
- Synthesize and distribute information from a range of sources to
present a more complete picture of the sources, distribution, fates,
and effects of contaminants in the Estuary ecosystem.
back to contents
Data Usage
Data from this program are made available for scientific research, regulatory
purposes, and public awareness. The RMP currently produces an annual report
that includes all the data, a summary of results, and some interpretation.
Examples of how the data are used by the RMP follows:
Trends:
Seasonal, annual, and long-term patterns in contaminants found in the
Estuary.
Objectives and Guidelines:
Data are used by the RMP to evaluate achievement of various water, sediment,
and tissue quality guidelines.
Conventional Water and Sediment Parameters:
Conventional water and sediment parameters are evaluated to see how
they affect contaminant levels. For example, how does sediment grain-size
affect sediment PAH concentrations, or how does DOC in water affect
water pesticide concentrations?
Integrated Contaminant Measurements:
Bioaccumulation data may be used to determine time-averaged trends in
contaminant concentrations and for comparison with other trend data.
Principal Contacts
Table 2 presents the principal contact representatives,
their affiliation with the RMP, abbreviations used in this report, and
current phone numbers.
back to contents
3. Overview of Quality Assurance and Control in the RMP
Sample Collection, Preservation and Holding
Field personnel will strictly adhere to the RMP protocols to ensure the
collection of representative, uncontaminated water, sediment, and tissue
chemistry samples. Briefly, the key aspects of quality control associated
with chemistry sample collection are as follows:
- Field personnel will be thoroughly trained in the proper use of sample
collection gear and will be able to distinguish acceptable versus unacceptable
samples in accordance with pre-established criteria.
- Field personnel will be thoroughly trained to recognize and avoid
potential sources of sample contamination (e.g., engine exhaust, winch
wires, deck surfaces, ice used for cooling).
- Samplers and utensils which come in direct contact with the sample
will be made of non-contaminating materials (e.g., glass, high-quality
stainless steel and/or Teflon®) and will be thoroughly
cleaned between sampling stations.
- Sample containers will be pre-cleaned and of the recommended type.
back to
contents
Laboratory Operations
The QA/QC requirements presented in the following sections are intended
to provide a common foundation for each laboratorys protocols; the
resultant QA/QC data will enable an assessment of the comparability of
results generated by different laboratories and different analytical procedures.
It should be noted that the QA/QC requirements specified in this plan
represent the minimum requirements for any given analytical method.
The RMPs performance-based protocols for all analytical laboratories
consist of two basic elements:
- Initial demonstration of laboratory capability. Prior to the initial
analysis of samples, each laboratory will demonstrate proficiency in
several ways: written protocols for the analytical methods to be employed
for sample analysis will be submitted to the Program for review; method
detection limits (MDLs) for each analyte will be provided, including
the method used for determining MDLs; an initial calibration curve will
be established for all analytes, the calibration curve shall include
a calibration point set at 3 to 5 times the MDL and should include a
minimum of 5 calibration points for trace organics; acceptable performance
will be shown on known or blind reference material (see Laboratory Quality
Control Procedures, Initial Demonstration of Capability, p. 20.); and
long-term standard reference material results on reference material
with comparable analyte concentrations as those in RMP field samples
will be submitted.
- Ongoing demonstration of capability. Following a successful first
phase, the laboratory will demonstrate its continued capabilities in
several ways: participation in an on-going series of interlaboratory
comparison exercises, routine analysis of certified reference materials,
calibration checks, and analysis of laboratory reagent blanks and fortified
samples. (See Laboratory Quality Control Procedures, Ongoing Demonstration
of Capability, p. 22.)
The results for the various QA/QC samples will be reviewed by laboratory
personnel immediately following the analysis of each sample batch. These
results will then be used to determine when data quality criteria have
not been met, and corrective actions will be taken before processing a
subsequent sample batch. When data quality criteria are not met, specific
corrective actions are required before the analyses may proceed.
back to
contents
Information Management
Various data and information generated from the RMP are stored at SFEI.
The digital data generated from the sampling cruises arrive at SFEI in
various formats and are converted to standard RMP database format. After
final QA checks, the data are uploaded to the RMP database in Oracle®.
Data tables are generated from this database. The same database is also
accessible through SFEIs website (http://www.sfei.org).
Sample Tracking
RMP sample collection personnel have developed a comprehensive system
for recording sampling information in the field and tracking sample shipments.
This component is included in the RMP Field Operations Manual (http://www.sfei.org/rmp/docs/fom_1.html).
back
to contents
Data Reporting Requirements
As previously indicated, laboratory personnel will verify that the measurement
process was "in control" (i.e., all specified data quality criteria
were met or acceptable deviations explained) for each batch of samples
before proceeding with the analysis of a subsequent batch. In addition,
each laboratory will establish a system for detecting and reducing transcription
and/or calculation errors prior to reporting data.
Only data which have met data quality criteria, or data which have acceptable
deviations explained, will be submitted by the laboratory. When QA requirements
have not been met, the samples will be reanalyzed when possible. Only
the results of the reanalysis will be submitted, provided they are acceptable.
back to contents
4. Field Quality Assurance and Quality Control
Field Performance Measurements: Terminology
Following is a list of definitions of field performance measurements
that are frequently included in the sampling protocol. Some of these measurements
only need to be taken when an established procedure is changed, while
others need to be taken at various intervals throughout the sampling process.
- Source Solution Blanks: These account for any pre-existing contamination
in the water or preservatives used to prepare the sample containers
as well as the field or travel blanks.
- Bottle Blanks: These account for contamination in sampling containers,
in addition to any contamination due to the source solution.
- Travel Blanks: These account for contaminants introduced during the
transport process between the laboratory and field site, in addition
to any contamination from the source solution and container.
- Equipment Blank: These account for contamination introduced by the
field sampling equipment.
- Field Duplicates: These account for variability in the field and
laboratory.
- Field Blanks: These account for all of the above sources of contamination
that might be introduced to a sample as well as that which would be
due to the sampling equipment and the immediate field environment. Field
blanks are generated under actual field conditions and are subjected
to the same aspects of sample collection, field processing, preservation,
transport, and laboratory handling as the environmental samples. Field
blanks for sediment analyses generally consist of ultra pure sand. True
field blanks for biological tissue samples do not exist.
back
to contents
Field Performance Measurements Used by the RMP
Routine preparation, collection, and analysis of all the field samples
mentioned above would be redundant and inefficient. Since trace metals
in environmental water samples are orders of magnitude lower than in sediments
or tissues, the field QA/QC measures are much more rigorous for water
samples. Most QA/QC steps taken to minimize trace element sampling artifacts
are also applicable for the collection of trace organic samples.
Source solution blanks will be made with Milli-Q or Nanopure water (free
of trace organic and element contaminants), and trace-metal grade acids
will be used in all aspects of cleaning, storage, and analysis. The sample
bottles will be cleaned and stored filled (water containers only) with
acid solution. Contamination of these source solutions will be routinely
checked, and corrective steps taken whenever contamination of source solutions
are indicated.
Bottle blanks that were generated early on in the monitoring program
showed that the "trace-metal clean" polyethlene and Teflon®
bottles used for all three of the RMP samples are not a source of
trace element contamination. Certified trace-metal-free borosilicate glass
containers will be used for sediment samples, and measurements of bottle
blanks will be conducted for each lot.
Travel blanks are not routinely used for water, sediment, or tissue samples.
The possibility of contamination during the transport between the laboratory
and field site will be mitigated by the measures taken to keep the sample
bottles in an enclosed micro-environment. All water sample bottles will
be quadruple-bagged and kept inside a tightly closed plastic bucket. They
will be filled with a weak acid solution, so any metals leached from the
container will be kept in solution. This storage solution will be discarded
immediately prior to sampling, followed by five rinses with the sample.
The sample bottles will be removed from the plastic bags only in a class
100 clean laboratory, except during active sample. The bottles will always
be handled with polyethylene-gloved clean hands.
Equipment blanks for water samples will be collected periodically in
the laboratory by pumping Milli-Q water through the sample tubing connected
to a filter cartridge. The sampling equipment will consist of a dual-head
peristaltic pump which pumps water up through the inlet length of Teflon®
tubing connecting to C-flex tubing, and finally to the outlet length of
Teflon® tubing. The Teflon® and C-flex tubing
will be connected via polypropylene Y connector fittings. Filtered samples
will additionally pass through a 0.45 micrometer polycabonate filter cartridge
attached to the outlet end. The sample will be exposed to the interior
of the Teflon® and C-flex tubing, the Y fittings, and the
filter cartridge, all of which will have been rigorously cleaned with
ultra-pure reagents. Sediments will be collected with a van Veen grab
sampler. However, equipment blanks will not be taken. The sediment sampling
protocol is discussed further in the field blanks section. Since bivalves
will be hand collected, equipment blanks are not relevant for tissue samples.
Field duplicates will only be routinely collected for water samples.
Water will be filtered in duplicate so that evaluation of the sampling
system precision includes the filter cartridge. Short-term environmental
variability, most notably due to swift currents and non-homogeneous suspended
sediment loads will affect the sampling precision. Golden Gate station
(BC20) probably has the least variability, and will, therefore, usually
be included as a field duplicate. Two or three additional stations at
different locations of the Bay will also be collected in duplicate.
Since sediment concentrations in the Estuary vary spatially, a field
duplicate would be unable to separate natural variability from that introduced
by the sampling and analysis system. In 1994, triplicate samples were
taken at three RMP stations to assess within-station variability. Variability
was shown to be parameter-specific for trace elements with certain metals
exhibiting less than 3% variability between triplicates and others up
to 40% variability between triplicates.
Field duplicates in bivalve tissue samples will not be collected per
se. Between 40 and 100 bivalves are deployed at each site. They will be
hand-collected and later homogenized as a single sample. Two sub-samples
of fewer animals each, would assess variability in the animals rather
than assess precision in technique or environmental variability.
Field blanks for water will be generated under actual field conditions
and will be treated in the exact same manner as the environmental field
samples in both the field and laboratory. True field blanks are, however,
difficult to obtain because assessment of the monitoring vessels
aura of contamination at the time of sampling is not straight-forward.
True field blanks are not routinely collected by any worker in this field
and are not routinely reported in the literature. Collection of a field
blank by pumping the "source solution" (Milli-Q water) through
the system on deck does not adequately address the issues of potential
contamination of the water sample by the monitoring vessel since metals
are ubiquitous on boats. Therefore, a field blank merely measures contamination
of the sampling equipment, which is already accounted for, and perhaps
aerosol contamination, but it cannot sort out vessel contamination from
water contamination present without the vessel sitting in the source water.
Mitigation steps for this potential problem will be taken. To avoid aerosol
contamination the sample tubing inlet and outlet will be kept covered
until the engines are turned off, and the engine will remain off until
sampling is completed and the tubing inlet and outlet are once again covered.
To avoid possible contamination of the sample by the boat, the 15-20 foot
sampling pole will be extended over the windward side, oriented up-current
from the vessel and upwind from the equipment and personnel.
To get around the inability to collect a true field blank, the metal
concentrations of environmental water samples will be considered accurate
if they are oceanographically consistent (Boyle et al., 1981), and comparable
values are obtained by intercalibration studies (Patterson and Settle,
1976). These mitigation methods have been adopted by many workers in the
field following extensive experience (Bewers and Windom, 1981; Boyle et
al., 1981; Schaule and Patterson, 1981; Berman et al., 1983; Bruland et
al., 1985; Flegal and Stukas, 1987; Landing et al., 1995; Yeats et al.,
1995).
Samples approaching field blanks have been obtained for the RMP by collecting
relatively pristine oceanic water well beyond coastal influences, using
the same research vessel and sampling equipment as during a normal sampling
cruise. The field blank will not be collected during the cruise, because
of the extra time required to motor the boat beyond coastal influences.
Routine collection of these oceanic blanks will not be conducted due to
cost constraints.
For trace organic sampling, containers will be routinely checked for
contamination, and plastic material for storage, transport, and protection
of samples will be avoided. Only ultra-pure solvents will be used in the
preparation of the XAD resin and filters that capture the particulate
and dissolved fraction of the water samples. The XAD resin and filters
through which about 100 liters of water are pumped will remain enclosed
and inaccessible to aerial contamination. Tests on travel blanks of XAD
columns and of a solvent-extracted glass fiber filter have shown either
no measurable levels of analytes or levels one to two orders of magnitude
lower than field concentrations (Jarman, in prep).
Collection of true sediment field blanks is logistically difficult and
has been deemed unnecessary due to precautions taken that minimize contamination
of the samples. Sediment samples will be collected with a van Veen grab
sampler based on modified NOAA Status and Trends, Benthic Surveillance
Project methods (Lauenstein and Young, 1986; SFEI, 1997) All surfaces
of sediment sampling and processing instruments coming into contact with
the sample will be made of inert materials, such as Teflon®
or stainless steel coated with Dykon®, and will be thoroughly
cleaned prior to field use. Equipment will also be cleaned with Alkonox
detergent between stations and rinsed with hydrochloric acid, followed
by methanol, to avoid any carryover contamination from one station to
another. Sampling, compositing, and homogenization will be conducted on
board ship with gloved hands, and the homogenate will be placed into pre-cleaned
polyethylene or Teflon® containers for trace element analyses,
and into pre-cleaned certified glass jars with Teflon®-lined
lids for trace organic analyses. The homogenization bucket will always
be covered with aluminum foil during the collection of the sediment samples
to avoid sample contamination via aerial deposition.
Bivalves will be handled in the field according to established protocols
of the California State Mussel Watch Program designed to minimize sample
contamination. Bivalves destined for trace element analysis will be placed
in polyethylene ziploc bags, placed on dry ice, and kept frozen until
homogenization and analysis. Bivalves used for trace organic analysis
will be wrapped in aluminum foil.
back to contents
|
|
5. Laboratory Quality Assurance and Control
RMP Laboratory Requirements
The San Francisco Estuary Institute requires all Regional Monitoring Program
laboratories to demonstrate capability continuously through:
- Strict adherence to common QA/QC procedures.
- Routine analysis of certified reference materials (CRMs).
- Regular participation in an on-going series of interlaboratory comparison
exercises.
This is a "performance-based" approach for measurements of
low-level contaminant analyses, involving continuous laboratory evaluation
through the use of accuracy-based materials (e.g., CRMs), laboratory matrix
spikes, laboratory reagent blanks, calibration standards, laboratory-
and field-duplicated blind samples, and others as appropriate. The definition
and use of each of these types of quality control samples are explained
in later sections.
Quality control operates to make sure that data produced are satisfactory,
consistent, and dependable. Under the RMP performance-based chemistry
QA program, laboratories are not required to use a single, standard analytical
method for each type of analysis, but rather are free to choose the best
or most feasible method within the constraints of cost and equipment that
is suitable for meeting the RMPs data quality criteria (DQCs). The
RMP DQCs were developed based on the kinds of general management questions
that the environmental data are supposed to help answer. The RMP has developed
specific guidelines for measurement precision, accuracy, and levels of
detection that are reflected in sampling, handling, and analysis requirements
that can satisfy a large spectrum of potential management questions. Each
laboratory will, however, continuously demonstrate proficiency and data
comparability through routine analysis of accuracy-based performance evaluation
samples, split samples, and reference materials representing actual sample
matrices. No single analytical method has been officially approved for
low-level (i.e., low parts per quadrillion and parts per billion) analysis
of organic and inorganic contaminants in water or estuarine sediments.
Recommended methods for the RMP are those developed in various academic
research programs and those used in the NOAA NS&T Program (Lauenstein
et al., 1993).
All laboratories providing analytical support for chemical or biological
analyses will have the appropriate facilities to store, prepare, and process
samples, and appropriate instrumentation and staff to provide data of the
required quality within the time period dictated by the project. Laboratories
are expected to conduct operations in a way that includes:
- A program of scheduled maintenance of analytical balances, microscopes,
laboratory equipment, and instrumentation.
- Routine checking of analytical balances using a set of standard reference
weights (American Society of Testing and Materials (ASTM) Class 3, NIST
Class S-1, or equivalents).
- Checking and recording the composition of fresh calibration standards
against the previous lot. Acceptable comparisons are < 2 percent
of the previous value.
- Recording all analytical data in bound (where possible) logbooks,
with all entries in ink, or electronic format.
- Monitoring and documenting the temperatures of cold storage areas
and freezer units once per week.
- Verifying the efficiency of fume hoods.
- Having a source of reagent water meeting ASTM Type I specifications
(ASTM, 1984) available in sufficient quantity to support analytical
operations. The conductivity of the reagent water will not exceed 18
megaohm at 25°C. Alternately, the resistivity of the reagent water
will exceed 10 m mhos/cm.
- Labeling all containers used in the laboratory with date prepared,
contents, initials of the individual who prepared the contents, and
other information as appropriate.
- Dating and safely storing all chemicals upon receipt. Proper disposal
of chemicals when the expiration date has passed.
- Having QAPP, SOPs, analytical methods manuals, and safety plans readily
available to staff.
- Having raw analytical data, such as chromatograms, accessible so
that they are available upon request.
Laboratories will be able to provide information documenting their ability
to conduct the analyses with the required level of data quality. Such
information might include results from interlaboratory comparison studies,
control charts and summary data of internal QA/QC checks, and results
from certified reference material analyses.
back to contents
Data Formatting and Transfer
Laboratories will also be able to provide analytical data and associated
QA/QC information in a format and time frame agreed upon with the RMP
Project Manager or designee. Each year data formatting and reporting expectations
will be clearly identified and distributed to participating laboratories.
Laboratory Personnel, Training, and Safety
Each laboratory providing analytical support to the RMP must have a designated
on-site QC Officer for the particular analytical component(s) performed
at that laboratory. This individual will serve as the point of contact
for the RMP QA staff in identifying and resolving issues related to data
quality.
To ensure that the samples are analyzed in a consistent manner throughout
the duration of the project, key laboratory personnel will participate
in an orientation session conducted during an initial site visit or via
communications with RMP staff. The purpose of the orientation session
is to familiarize key laboratory personnel with the QAPP and the QA/QC
program. Participating laboratories may be required to demonstrate acceptable
performance before analysis of samples can proceed, as described in subsequent
sections. Laboratory operations will be evaluated on a continuous basis
through technical systems audits, and by participation in interlaboratory,
round-robin programs. Meetings shall be held with all participating laboratories
at regular intervals to continually review QA/QC procedures, and to revise/update
the QAPP.
Personnel in any laboratory performing RMP analyses will be well versed
in good laboratory practices, including standard safety procedures. It
is the responsibility of the particular analytical component project officer,
laboratory manager, and/or supervisor to ensure that safety training is
mandatory for all laboratory personnel. Each laboratory is responsible
for maintaining a current safety manual in compliance with the Occupational
Safety and Health Administration (OSHA), or equivalent state or local
regulations. The safety manual will be readily available to laboratory
personnel. Proper procedures for safe storage, handling, and disposal
of chemicals will be followed at all times; each chemical will be treated
as a potential health hazard and good laboratory practices will be implemented
accordingly.
back to contents
Quality Assurance Documentation
All laboratories will have the latest revision of the RMP QAPP. In addition,
the following documents and information will be current, and they will be
available to all laboratory personnel participating in the processing of
RMP samples, as well as to SFEI project officials:
- Laboratory QA Plan: Clearly defined policies and protocols specific
to a particular laboratory, including personnel responsibilities, laboratory
acceptance criteria and corrective actions to be applied to the affected
analytical batches, qualification of data, and procedures for determining
the acceptability of results.
- Laboratory Standard Operating Procedures (SOPs): Containing instructions
for performing routine laboratory procedures.
- Laboratory Analytical Methods Manual: Step-by-step instructions describing
exactly how a method is implemented in the laboratory for a particular
analytical procedure. Contains all analytical methods utilized in the
particular laboratory for the RMP.
- Instrument Performance Information: Information on instrument baseline
noise, calibration standard response, analytical precision and bias
data, detection limits, etc. This information is usually recorded in
logbooks or laboratory notebooks.
- Control Charts: Control charts are useful in evaluating internal
laboratory procedures and are helpful in identifying and correcting
systematic error sources. Contract laboratories are encouraged to develop
and maintain control charts whenever they may serve in determining sources
of analytical problems.
back to contents
Laboratory Performance Audits/Corrective Action
Initially, a QA performance audit will be performed by RMP QA staff to
determine if each laboratory effort is in compliance with the procedures
outlined in the QAPP and to assist the laboratory where needed. Additionally,
technical systems audits will be conducted by a team composed of the RMP
QA Officer or designee, and his/her technical assistants. Reviews may
be conducted at any time during the scope of the study. Results will be
reviewed with participating laboratory staff and corrective action recommended
and implemented, where necessary. Furthermore, laboratory performance
will be assessed on a continuous basis through the use of laboratory intercomparison
studies (round robins). Laboratories performing organic and metal chemistry
analyses will be required to participate in the annual National Status
and Trends Intercalibration, and to report the findings in a timely fashion
to the designated contact at NOAA and to the RMP QA Officer.
back
to contents
Laboratory Performance Measurements
Laboratory performance measurements included in the analysis stream and
are designed to check if data quality criteria are met are briefly defined
below.
- Method Blanks (also called laboratory reagent blanks or preparation
blanks): These account for contaminants present in the preservative
and analytical solutions used during the quantification of the parameter.
- Injection Internal Standards: This accounts for error introduced by
the analytical instrument.
-
- Replicate Samples: These are replicates of extracted material
that measure the instrumental precision.
- Laboratory Replicate Samples: These are replicates of the raw
material that are extracted and analyzed to measure laboratory precision.
- Matrix Spike Replicate Samples: This is used to assess both laboratory
precision and accuracy. This is particularly useful when the field
samples analyzed do not contain many of the target compounds (measuring
non-detects in replicate does not allow the data reviewer to measure
the precision or the accuracy of the data in an analytical batch).
- Matrix Spike Samples: These are field samples to which a known
amount of contaminant is added and used to measure potential analytical
interferences present in the field sample.
- Certified Reference Materials (CRM): Analysis of CRMs is another way
of determining accuracy of the analysis by comparing a certified value
of material with similar concentrations as those expected in the samples
to be analyzed.
These types of samples serve to check if errors were introduced during
the analysis process and if so, at what step(s) and at what magnitude.
The remainder of this document will provide RMP guidance for general laboratory
requirements, and protocols for checking and tracking possible sources
of errors (outlined above) in the analytical process.
back
to contents
Laboratory Quality Control Procedures
The performance-based protocols utilized in the RMP for analytical chemistry
laboratories consist of two basic elements: initial demonstration of laboratory
capability (e.g., documentation that the analyses of samples are within
the data quality criteria) and ongoing demonstration of capability. Prior
to the initial analysis of samples, each laboratory will demonstrate capability
and proficiency.
Initial Demonstration of Capability
Instrument Calibration
Upon initiation of an analytical run, after each major equipment disruption,
and whenever on-going calibration checks do not meet recommended DQCs
(see Tables 3 and 4), the
system will be calibrated with a full range of analytical standards. Immediately
after this procedure, the initial calibration must be verified through
the analysis of a standard obtained from a different source than the standards
used to calibrate the instrumentation and prepared in an independent manner
and ideally having certified concentrations of target analytes of a certified
reference material (CRM) or certified solution. Frequently, calibration
standards are included as part of an analytical run, interspersed with
actual samples. However, this practice does not document the stability
of the calibration and is incapable of detecting degradation of individual
components, particularly pesticides, in standard solutions used to calibrate
the instrument. The calibration curve is acceptable if it has a r2
of 0.990 or greater for all analytes present in the calibration mixtures.
If not, the calibration standards, as well as all the samples in the batch
must be re-analyzed. All calibration standards will be traceable to a
recognized organization for the preparation and certification of QA/QC
materials (e.g., NIST, National Research Council Canada (NRCC), US EPA,
etc.).
Calibration curves will be established for each analyte and batch analysis
from a calibration blank and a minimum of three analytical standards of
increasing concentration, covering the range of expected sample concentrations.
Only data which result from quantification within the demonstrated working
calibration range may be reported by the laboratory (i.e., quantification
based on extrapolation is not acceptable). Alternatively, if the instrumentation
is linear over the concentration ranges to be measured in the samples,
the use of a calibration blank and one single standard that is higher
in concentration than the samples may be appropriate. Samples outside
the calibration range will be diluted or concentrated, as appropriate,
and reanalyzed.
back
to contents
Initial Documentation of Method Detection Limits
Analytical chemists have coined a variety of terms to define "limits"
of detectability; definitions for some of the more commonly used terms
are provided in Keith et al. (1983) and in Keith (1991). In the RMP, the
method detection limit (MDL) is used to define the analytical limit of
detectability. The MDL represents a quantitative estimate of low-level
response detected at the maximum sensitivity of a method. The Code of
Federal Regulations (40 CFR Part 136) gives the following rigorous definition:
The MDL is the minimum concentration of a substance that can be measured
and reported with 99% confidence that the analyte concentration is greater
than zero and is determined from analysis of a sample in a given matrix
containing the analyte.
The American Society of Testing and Materials (ASTM) defines the limit
of detection as:
A concentration of twice the criterion of detection...when it has been
decided that the risk of making a Type II error is to be equal to a Type
I error.
In order to compare MDLs in quantitative terms by different laboratories
participating in RMP analysis, MDLs will initially be determined according
to 40 CFR 136.2 (f) and Appendix B of 40 CFR 136. Determining the MDL with
this procedure is elaborate and need not be determined annually provided
that:
- No process or method changes have been made.
- Check samples containing an analyte spike at about 2x MDL indicate
that the sample is detected. The required frequency of check samples
is quarterly.
The matrix and the amount of sample (i.e., dry weight of sediment or
tissue) used in calculating the MDL will match as closely as possible
the matrix of the actual field samples and the amount of sample typically
used. In order to ensure comparability of results among different laboratories,
MDL target values have been established for the RMP (see Table
5). These MDLs have been derived empirically. Most are considerably
lower than water quality objectives or sediment and tissue quality guidelines
and provide the foundation for having a high level of certainty in the
data.
The laboratory shall confirm the ability to analyze low-level samples
with each batch. This shall be accomplished by analyzing a method blank
spiked at 3 to 5 times the method detection limit. Recoveries for organic
analyses shall be between 50 and 150% for at least 90% of the target analytes.
back to contents
Limits of Quantitation
Taylor (1987) states that "a measured value becomes believable when
it is larger than the uncertainty associated with it". The uncertainty
associated with a measurement is calculated from the standard deviation
of replicate measurements (s0) of a low concentration standard
or a blank. Normally, the MDL is set at three times the standard deviation
of replicate measurements, as it is at this point that the uncertainty
of a measurement is approximately ±100% at the 95% level of confidence.
Values at the MDL may not reflect a signal much above zero and, therefore,
are quantitatively not very meaningful. The limit of quantitation (LOQ),
as established by the American Chemical Society, is normally ten times
the standard deviation of replicate measurements, which corresponds to
a measurement uncertainty of ±30% (see Taylor, 1987). By these standard
definitions, measurements below the MDL are not believable, measurements
between the LOQ and the MDL are only semi-quantitative, and confidence
in measurements above the LOQ is high.
back to contents
Initial Blind Analysis of Representative Sample
As appropriate, representative sample matrices which are uncompromised,
homogeneous, and contain the analytes of interest at concentrations of
interest will be used to evaluate performance of analytical laboratories
new to the RMP prior to the analysis of field samples. The samples used
for this initial demonstration of laboratory capability typically will
be distributed blind (i.e., the laboratory will not know the concentrations
of the analytes of interest) as part of the interlaboratory comparison
exercises. Based on results that have typically been attained by experienced
RMP laboratories, a new laboratorys performance generally will be
considered acceptable if its submitted values are within DQCs (Tables
3 and 4) of the known concentration, or the consensus value, of each analyte
of interest in the samples. These criteria apply only for analyte concentrations
equal to or greater than three times the RMP target MDL. If the results
for the initial analysis fail to meet these criteria, the laboratory will
be required to repeat the analysis until the performance criteria are
met, prior to the analysis of RMP field samples.
back to contents
Record of Certified Reference Material
As CRMs are routinely included in analysis of batches of reputable laboratories,
the historical record of results may also serve as a suitable performance
indicator.
Ongoing Demonstration of Capability
Participation in Interlaboratory Comparison Exercises
Through an interagency agreement, NOAAs NS&T Program and EPAs
EMAP program jointly sponsor an on-going series of interlaboratory comparison
exercises (round-robins). All the RMP analytical laboratories are required
to participate in these intercomparison exercises, which are conducted
jointly by NIST and NRCC. These exercises provide a tool for continuous
improvement of laboratory measurements by helping analysts identify and
resolve problems in methodology and/or QA/QC. The results of these exercises
are also used to evaluate both the individual and collective performance
of the participating analytical laboratories on a continuing basis and
to insure that ongoing measurements are meeting DQCs. The RMP laboratories
are required to initiate corrective actions if their performance in these
comparison exercises falls below certain pre-determined minimal standards,
described in later sections.
One exercise is usually conducted over the course of a year. In a typical
exercise, NIST or NRCC will distribute performance evaluation samples
of an "unknown" and a certified reference material (CRM) to
each laboratory, along with detailed instructions for analysis. A variety
of performance evaluation samples have been utilized in the past, including
accuracy-based solutions, sample extracts, and representative matrices
(e.g., sediment or tissue samples). Laboratories are required to analyze
the sample(s) "blind" and will submit their results in a timely
manner both to the RMP Coordinator and to NIST or NRCC (as instructed).
Laboratories which fail to maintain acceptable performance may be required
to provide an explanation and/or undertake appropriate corrective actions.
At the end of each calendar year, coordinating personnel at NIST and NRCC
hold a QA workshop to present and discuss the comparison exercise results.
Representatives from participating laboratories are strongly encouraged
to participate in the annual QA workshops, which provide a forum for discussion
of analytical problems brought to light in the comparison exercises.
back to contents
Routine Analysis of Certified Reference Materials or Laboratory Control
Materials
Certified reference materials generally are considered the most useful
QC samples for assessing the accuracy of a given analysis (i.e., the closeness
of a measurement to the "true" value). CRMs can be used to assess
accuracy because they have "certified" concentrations of the
analytes of interest, as determined through replicate analyses by a reputable
certifying agency using two independent measurement techniques for verification.
In addition, the certifying agency may provide "non-certified"
or "informational" values for other analytes of interest. Such
values are determined using a single measurement technique, which may
introduce unrecognized bias. Therefore, non-certified values must be used
with caution in evaluating the performance of a laboratory using a method
which differs from the one used by the certifying agency.
A laboratory control material (LCM) is similar to a certified reference
material in that it is a homogeneous matrix which closely matches the
samples being analyzed. A "true" LCM is one which is prepared
(i.e., collected, homogenized, and stored in a stable condition) strictly
for use in-house by a single laboratory. Alternately, the material may
be prepared by a central laboratory and distributed to others (so-called
regional or program control materials). Unlike CRMs, concentrations of
the analytes of interest in LCMs are not certified but are based upon
a statistically valid number of replicate analyses by one or several laboratories.
In practice, this material can be used to assess the precision (i.e.,
consistency) of a single laboratory, as well as to determine the degree
of comparability among different laboratories. If available, LCMs may
be preferred for routine (i.e., day to day) analysis because CRMs are
relatively expensive.
Routine analysis of CRMs or, when available, LCMs represents a particularly
vital aspect of the "performance-based" RMP QA philosophy. At
least one CRM or LCM must be analyzed along with each batch of 20 or fewer
samples (i.e., QA samples should comprise a minimum of 5% of each set
of field samples). For CRMs, both the certified and non-certified concentrations
of the target analytes will be known to the analyst(s) and will be used
to provide an immediate check on performance before proceeding with a
subsequent sample batch. Performance criteria for both precision and accuracy
have been established for analysis of CRMs or LCMs (Tables 3 and 4); these
criteria are discussed in detail in the following paragraphs. If the laboratory
fails to meet either the precision or accuracy control limit criteria
for a given analysis of the CRM or LCM, the data for the entire batch
of samples is suspect. Calculations and instruments will be checked; the
CRM or LCM may have to be reanalyzed (i.e., reinjected) to confirm the
results. If the values are still outside the control limits in the repeat
analysis, the laboratory is required to find and eliminate the source(s)
of the problem and repeat the analysis of that batch of samples until
control limits are met, before final data are reported. The results of
the CRM or LCM analysis will never be used by the laboratory to "correct"
the data for a given sample batch.
Precision criteria: Precision is the reproducibility of an analytical
method. Each laboratory is expected to maintain control charts for use
by analysts in monitoring the overall precision of the CRM or LCM. Upper
and lower control chart limits (e.g., warning limits and control limits)
will be continually updated; control limits based on 99% confidence intervals
around the mean are recommended. The relative standard deviation (RSD)
will be calculated for each analyte of interest in the CRM based on the
last 7 CRM analyses. Acceptable precision targets for various analyses
are listed in Tables 3 and 4.
back to contents
Laboratory Replicates for Precision
A minimum of one field sample per batch of RMP samples submitted to the
laboratory will be processed and analyzed in duplicate or more for precision.
The relative percent difference between two replicate samples or the relative
standard deviation between more than two replicate samples (RPD or RSD
respectively) will be less than the DQC listed in Tables 3 and 4 for each
analyte of interest. Following are the calculations:

If results for any analytes do not meet the DQC for the RPD or RSD, calculations
and instruments will be checked. A repeat analysis may be required to
confirm the results. Results which repeatedly fail to meet the objectives
indicate sample inhomogeneity, unusually high concentrations of analytes
or poor laboratory precision. In this case, the laboratory is obligated
to halt the analysis of samples and eliminate the source of the imprecision
before proceeding.
Accuracy criteria: The "absolute" accuracy of an analytical
method can be assessed using CRMs only when certified values are provided
for the analytes of interest. However, the concentrations of many analytes
of interest to the RMP are provided only as non-certified values in some
of the more commonly used CRMs. Therefore, control limit criteria are
based on "relative accuracy", which is evaluated for each analysis
of the CRM or LCM by comparison of a given laboratorys values relative
to the "true" or "accepted" values in the LCM or CRM.
In the case of CRMs, this includes both certified and noncertified values.
The "true" values are defined as the 95% confidence intervals
of the mean.
Based on typical results attained by experienced analysts in the past,
accuracy control limits have been established both for individual compounds
and combined groups of compounds (Tables 3 and 4).
There are three combined groups of compounds for the purpose of evaluating
relative accuracy for organic analyses: PAHs, PCBs, and pesticides. For
each group of analytes, 70% of the individual analytes will be within
35% of the certified 95% confidence interval; no individual analyte value
shall exceed ±30% of the 95% confidence interval more than once in consecutive
analyses without appropriate documentation and consultation with the RMP
QA officer. For inorganic analyses, the laboratorys value will be
within 2025% of the certified 95% confidence interval for each analyte
of interest in the CRM. Due to the inherent variability in analyses near
the method detection limit, control limit criteria for relative accuracy
only apply to analytes with true values which are >3 times the MDL
established by the laboratory.
back to contents
Continuing Calibration Checks
Calibration check solutions traceable to a recognized organization must
be inserted as part of the sample stream. The source of the calibration
check solution shall be independent from the standards used for the calibration.
Calibration check solutions used for the continuing calibration checks
will contain all the analytes of interest. The frequency of these checks
is dependent on the type of instrumentation used and, therefore, requires
considerable professional judgment. All organic analyses shall be bracketed
by an acceptable calibration check. A calibration check standard shall
be run every 12 hours at a minimum.
If the control limits for analysis of the calibration check solution
(set by the laboratories) are not met, the initial calibration will have
to be repeated. The calibration check for 90% of the analyte shall not
deviate more than ±25% from the known value for PAHs and ±20% for PCBs
and pesticides. If possible, the samples analyzed before the calibration
check solution that failed the DQCs will be reanalyzed following recalibration.
The laboratory will begin by reanalyzing the last sample analyzed before
the calibration check solution which failed. If the RPD between the results
of this reanalysis and the original analysis exceeds precision DQCs (Tables
3 and 4), the instrument is assumed to have been out of control during
the original analysis. If possible, reanalysis of samples will progress
in reverse order until it is determined that the RPD between initial and
reanalysis results are within DQCs (Tables 3 and 4). Only the re-analysis
results will be reported by the laboratory. If it is not possible or feasible
to perform reanalysis of samples, all earlier data (i.e., since the last
successful calibration control check) are suspect. In this case, the laboratory
will prepare a narrative explanation to accompany the submitted data.
back to contents
Laboratory Reagent Blank
Laboratory reagent blanks (also called method blanks, extraction blanks,
procedural blanks, or preparation blanks) are used to assess laboratory
contamination during all stages of sample preparation and analysis. For
both organic and inorganic analyses, one laboratory reagent blank will
be run in every sample batch. The reagent blank will be processed through
the entire analytical procedure in a manner identical to the samples.
Reagent blanks should be less than the MDL or not exceed a concentration
greater than 10% of the lowest reported sample concentration. A reagent
blank concentration > 2x the MDL or > 10% of the lowest reported
sample concentration for one or more of the analytes of interest will
require corrective action to identify and eliminate the source(s) of contamination
before proceeding with sample analysis.
If eliminating the blank contamination is not possible, all impacted
analytes in the analytical batch shall be flagged. In addition, a detailed
description of the contamination source and the steps taken to eliminate/minimize
the contaminants shall be included in the transmittal letter. Subtracting
method blank results from sample results is not permitted.
back to contents
Completeness
Completeness is defined as "a measure of the amount of data collected
from a measurement process compared to the amount that was expected to
be obtained under the conditions of measurement" (Stanley and Verner,
1985). Field personnel will always strive to achieve or exceed the RMP
completeness goals of 9598% for water, sediment, or tissue samples.
Surrogates
The usage of the terms "surrogate", "injection internal
standard", and "internal standard" varies considerably
among laboratories and is clarified here.
Surrogates are compounds chosen to simulate the analytes of interest
in organic analyses. Surrogates are used to estimate analyte losses
during the extraction and clean-up process and must be added to each sample,
including QA/QC samples, prior to extraction. The reported concentration
of each analyte is adjusted to correct for the recovery of the surrogate
compound, as done in the NOAA NS&T Program. The surrogate recovery
data will be carefully monitored; each laboratory must report the percent
recovery of the surrogate(s) along with the target analyte data for each
sample. If possible, isotopically-labeled analogs of the analytes will
be used as surrogates.
Each laboratory will set its own warning limit criteria based on the
experience and best professional judgment of the analyst(s). It is the
responsibility of the analyst(s) to demonstrate that the analytical process
is always "in control" (i.e., highly variable surrogate recoveries
are not acceptable for repeat analyses of the same certified reference
material and for the matrix spike/matrix spike duplicate). The warning
limit criteria used by the laboratory will be provided in the standard
operating procedures submitted to SFEI.
back to contents
Internal Standards
For gas chromatography (GC) analysis, internal standards (also referred
to as "injection internal standards" by some analysts) are added
to each sample extract just prior to injection to enable optimal quantification,
particularly of complex extracts subject to retention time shifts relative
to the analysis of standards. Internal standards are essential if the
actual recovery of the surrogates added prior to extraction is to be calculated.
The internal standards can also be used to detect and correct for problems
in the GC injection port or other parts of the instrument. The compounds
used as internal standards will be different from those already used as
surrogates. The analyst(s) will monitor internal standard retention times
and recoveries to determine if instrument maintenance or repair, or changes
in analytical procedures, are indicated. Corrective action will be initiated
based on the judgment of the analyst(s). Instrument problems that may
have affected the data or resulted in the reanalysis of the sample will
be documented properly in logbooks and internal data reports and used
by the laboratory personnel to take appropriate corrective action.
Dual-Column Confirmation
Dual-column chromatography is required for analyses using GC-ECD due
to the high probability of false positives arising from single-column
analyses.
back to contents
Matrix Spike and Matrix Spike Duplicate
A laboratory fortified sample matrix (commonly called a matrix spike,
or MS) and a laboratory fortified sample matrix duplicate (commonly called
a matrix spike duplicate, or MSD) will be used both to evaluate the effect
of the sample matrix on the recovery of the compound(s) of interest and
to provide an estimate of analytical precision. A minimum of 5% of the
total number of samples submitted to the laboratory in a given year will
be selected at random for analysis as matrix spikes/matrix spike duplicates.
A field sample is first homogenized and then split into three subsamples.
Two of these subsamples are fortified with the matrix spike solution and
the third subsample is analyzed to provide a background concentration
for each analyte of interest. The matrix spike solution should contain
as many representative analytes from the RMP analyte list as feasible.
The final spiked concentration of each analyte in the sample will be at
least 10 times the MDL for that analyte, as previously calculated by the
laboratory. Additionally, the total number of spikes should cover the
range of expected concentrations. Recovery is the accuracy of an analytical
test measured against a known analyte addition to a sample. Recovery is
calculated as follows:
Recovery data for the fortified compounds ultimately will provide a basis
for determining the prevalence of matrix effects in the samples analyzed
during the project. If the percent recovery for any analyte in the MS
or MSD is less than the recommended warning limit of 50 percent, the chromatograms
(in the case of trace organic analyses) and raw data quantitation reports
will be reviewed. If an explanation for a low percent recovery value is
not discovered, the instrument response may be checked using a calibration
standard. Low matrix spike recoveries may be a result of matrix interferences
and further instrument response checks may not be warranted, especially
if the low recovery occurs in both the MS and MSD, and the other QC samples
in the batch indicate that the analysis was "in control". An
explanation for low percent recovery values for MS/MSD results will be
discussed in a cover letter accompanying the data package. Corrective
actions taken and verification of acceptable instrument response will
be included. Analysis of the MS/MSD is also useful for assessing laboratory
precision. The RPD between the MS and MSD results should be less than
the target criterion listed in Tables 3 and 4 for each analyte of interest.
back to contents
Field Replicates and Field Split Samples
As part of the regular quality assurance program of the RMP, replicate
sediment and tissue samples may be collected, homogenized, and placed
in separate sample containers at a minimum of one pre-selected station
for subsequent chemical analysis whenever funds allow. One of the sample
containers for each trace organic and metals analysis will be submitted
as a blind field replicate to the primary analytical laboratory. Another
set of containers, called field splits, will be sent blind to additional
laboratories selected to participate in the split sample analysis of trace
elements and trace organics. The analysis of field replicates and field
splits will provide an assessment of both inter-and intra-laboratory precision
and variance in the sample matrix at the field site.
|
|
|
QA Procedures for Ancillary Parameters in Water, Sediment Toxicity, Bivalve
Condition, and Fish Tissue
Several ancillary parameters are measured in water and sediment.
Water
Toxicity
There will be five replicates per sample, plus three for water quality.
Test containers will be glass scintillation vials with 10 ml test solution.
Organisms and samples will be maintained at appropriate temperatures.
All instruments will be properly calibrated. Toxicity test procedures
are considered unacceptable if the percentage of normal live larvae is
below 70% of test controls. Acceptable temperatures range from 14°
to 16°C; acceptable salinities range from 26 to 30; acceptable
dissolved oxygen concentrations range from 5.09 to 8.49 mg/L.
DOC
Blanks will be analyzed a minimum of three times each day during sample
analysis. The instrument will be calibrated with a standard curve at least
once every 15 samples. Duplicate field samples will be obtained and analyzed
from every station, with a minimum of three measurements being made on
each field duplicate. The reported values are the averages of the six
measurements made on the two duplicates from each stations. Although no
standard for DOC in water is commercially available, an internal laboratory
reference material will be analyzed a minimum of three times during sample
analysis. The criteria for both precision and accuracy is ±5%.
back to contents
TSS
The analytical balance used in the gravimetric measurement of TSS has
an internal checking device and will be periodically checked by a service
representative. A minimum of three blanks are analyzed during sample analysis.
As sample volume permits, samples from approximately three stations will
be analyzed in duplicate or triplicate. No standard is available for TSS.
Precision is ±5%.
Chlorophyll
The fluorometer used to measure chlorophyll and phaeophytin will be calibrated
twice annually using a chlorophyll standard that has been analyzed by
UV-VIS spectrometry. A blank will be analyzed with the samples. Duplicate
filtrates will be obtained in the field for each station and each filter
will be analyzed at least once. The reported values are the averages of
the measurements for the duplicate filtrates. The precision criterion
is ±10%.
Nutrients
The spectrometer used to analyze nutrients (i.e., ammonia, nitrate, nitrite,
phosphates, and silicates) will be calibrated with a standard curve based
on dilutions of stock standards that are mixed fresh for the analysis
of each cruise. Three blanks will be analyzed with each nutrient. Duplicate
aliquots will be analyzed from the field sample for each station. As sample
volume permits, at least one station will be analyzed in triplicate. The
SPEC QCS reference material of nutrients in wastewater will be analyzed
once during analyses for each cruise, although it contains reported concentrations
of only ammonia, nitrate, and phosphate. There are no commercially-available
reference materials for silicate and nitrite. The precision criterion
is ±5% and accuracy criterion is ±10%.
back to contents
Salinity
The salinometer used to analyze salinity will be calibrated annually
with IAPSO Standard Seawater reference material. A minimum of two blanks
will be analyzed during sample analysis. All stations will be analyzed
twice and the reported values are the average of the measurements for
each station. The precision criterion is ±1%.
CTD
The CTD will be returned to the manufacturer annually for recalibration
of all probes. The resulting revised calibration constants will then be
entered in a configuration file in SeaSoft (v. 4.035b) that is named corresponding
to its date of implementation so that the appropriate configuration file
can always be applied to any data set.
back to contents
Sediment
Bioassays
There will be five replicates per sample, plus a sixth for water quality.
Test containers will be glass for sediments and plastic for the reference
toxicant. Organisms and samples will be maintained at appropriate temperatures.
All instruments will be calibrated properly. Toxicity test procedures
are considered unacceptable if amphipod survival in home sediment controls
is less than 90%, or if survival in any control replicate is less than
80%. Acceptable temperature range is from 14° to 16°C, acceptable
salinities range from 17 to 23, acceptable dissolved oxygen
concentrations range from 5.09 to 8.49 mg/L.
TOC
Blanks and a reference material supplied by the instrument manufacturer,
Coulometrics, Inc. will be analyzed a minimum of three times daily during
sample analysis. The precision criterion is ±3% and accuracy criterion
is ±1%.
Grain Size
Standard reference materials will be analyzed with every batch of samples.
These include NIST SRM 1003b glass spheres and a narrow-sized garnet standard
supplied by the instrument manufacturer. In addition, at least one sample
in twelve will be analyzed in duplicate to determine precision. The precision
criterion is ±20%.
back to contents
Porewater Ammonia
The calibration of the ammonia probe on the pH/ORP meter will be checked
before analysis of each station. The calibration curve will also be used
for quantification of ammonia from millivolt potential readings made in
the samples with the ammonia probe. The calibration curve will be made
with reference standards of 10.0, 5.0, 1.0, and 0.1 ppm total ammonia
using dilutions of a NIST-traceable 1,000-ppm standard (Corning #951007).
New reference standards will be prepared and the probe will be recalibrated
if the millivolt reading for a particular standard drifts by more than
10% from the original reading. During sample analysis the probe is allowed
to remain in the sample until stable readings are achieved and recorded.
Porewater pH
Calibration of the pH probe on the pH/ORP meter will be performed before
sampling each station using reference standards of 4.0, 7.0, and 10.0
pH acidity. The standards will be made before each cruise from NIST-traceable
materials.
back to contents
Bivalves and Fish Tissue
Bivalve Condition Index
The precision of displacement volume measurements will be estimated by
making 10 separate measurements on a single organism.
Butyltins
Assessment of the distribution and environmental impact of butyltins
require measurements in marine sediment and tissue samples at trace levels.
Quality control of these measurements consists of checks on laboratory
precision and accuracy. One laboratory reagent blank must be run with
each batch of 25 or fewer samples. A reagent blank concentration between
the MDL and 3 times the MDL will serve as a warning limit requiring further
investigation based on the best judgment of the analyst(s). A reagent
blank concentration equal to or greater than 3 times the MDL requires
corrective action to identify and eliminate the source(s) of contamination,
followed by re-extraction and reanalysis of the samples in the associated
batch.
One laboratory fortified sample matrix (commonly called a matrix spike)
or laboratory fortified blank (i.e., spiked blank) will be analyzed along
with each batch of 25 or fewer samples to evaluate the recovery of the
butyltin species of interest, if authorized and funded. The butyltins
will be added at 5 to 10 times the MDLs as previously calculated by the
laboratory. If the percent recovery for any of the butyltins in the matrix
spike or spiked blank is outside the range 70 to 130 percent, analysis
of subsequent sample batches will stop until the source of the discrepancy
is determined and the system corrected.
Lipids
Lipid measurements are essential to interpretation of temporal or spatial
trends in concentrations of organic contaminants in tissues. Data quality
criteria for precision will apply to analysis of SRMs and laboratory duplicates.
For repeated analysis of SRMS, RPD should be <35% or RSD should be
<30%. For laboratory duplicates, RPD should be <35%.
back to contents
|
|
|
References
Berman, S.S., R.E. Sturgeon, J.A.H. Desaulniers, and A.P. Mykytiuk.
1983. Preparation of the sea water reference material for trace metals,
NASS-1. Marine Pollution Bulletin 14:69-73.
Bewers, J.M. and H.L. Windom. 1981. Intercomparison of seawater sampling
devices for trace metals. In Trace Metals in Seawater, C.S. Wong, E.
Boyle, K.W. Bruland, J.D. Burton, and E.D. Goldberg, eds. Plenum Press,
New York, pp. 143-154.
Boyle, E.A., S.S. Huested, and S.P. Jones. 1981. On the distribution
of copper, nickel, and cadmium in the surface waters of the North Atlantic
and North Pacific Ocean. Journal of Geophysical Research 86:9:8048-8066.
Bruland, K.W., K.H. Coale, and L. Mart. 1985. Analysis of seawater
for dissolved cadmium, copper, and lead: an intercomparison of voltametric
and atomic absorption methods. Mar. Chem. 17:285-300.
Cantillo, A.Y. and J.A. Calder. 1992. Reference materials for marine
science. Fresenius Journal of Analalytic Chemistry 338(4):380-382.
Flegal, A.R., and V.J. Stukas. 1987. Accuracy and precision of lead
isotopic composition measurements in seawater. Marine Chemistry 22:163-177.
Keith L.H. 1991. Environmental Sampoing and Analysis: A Practical Guide.
Lewis Publishers, Chelsea, MI, 143 pp.
Keith et al., 1983. Principles of Environmental Analysis. Anal. Chem.
55:2210-2218.
Landing, W.M., G.A. Cutter, J.A. Dalziel, A.R. Flegal, R.T. Powell,
D. Schmidt, A. Shiller, P. Statham, S. Westerlund, and J. Resing. 1995.
Analytical intercomparison results from the 1990 Intergovernmental Oceanographic
Commission open-ocean baseline survey for trace metals: Atlantic Ocean.
Marine Chemistry 49:253-265.
Lauenstein et al., July, 1993. NOAA Technical Memorandum NOS ORCA 71.
National Status and Trends Program for Marine Environmental Quality.
Silver Spring, Maryland.
Lauenstein, G.G., and D.R. Young. 1986. National Status and Trends
Program for Environmental Quality. Benthic Surveillance Project: Cycle
III Field Manual. NOAA Tech. Memo NOS OMA28.
Patterson, C.C. and D.M. Settle. 1976. The reduction of orders of magnitude
errors in lead analyses of biological materials and natural waters by
evaluating and controlling the extent and sources of industrial lead
contamination introduced during sample collecting, handling, and analysis.
In Accuracy in Trace Analysis: Sampling, Sample Handling, Analysis,
P. LaFleur, ed. National Bureau of Standards Special Publication 422:321-351.
Schaule, B.K. and C.C. Patterson. 1981. Lead concentrations in the
northeast Pacific: evidence for global anthropogenic perturbations.
Earth Planet Sci. Lett. 54:97-116.
SFEI. 1997. 1996 Annual Report: San Francisco Estuary Regional Monitoring
Program for Trace Substances. San Francisco Estuary Institute, Richmond,
CA.
SFEP. 1993. Comprehensive Conservation and Management Plan. San Francisco
Estuary Project, Oakland, CA.
Smith, D.R. and A.R. Flegal. 1993. Silver in San Francisco Bay estuarine
waters. Estuaries 16:547-558.
Stanley, T.W. and S.S. Verner 1985. The U.S. Environmental Progection
Agencys Quality Asurance Program. In: Quality Assurance for Environmental
Measurements, ASTM STP 867, J.K. Taylor and T.W. Stanley, Eds. American
Society for Testing and Materials, Philadelphia, PA, pp.12-19
Taylor J.K., 1987. Quality Assurance of Chemical Measurements. CRC
Press Inc.
Yeats, P.A., S. Westerlund, and A.R. Flegal. 1995. Cadmium, copper
and nickel distributions at four stations in the eastern central and
south Atlantic. Marine Chemistry 49:283-293.
back to contents
|
|